The PMA preempts. The 510(k) doesn't.

21 U.S.C. § 360k(a) bars a state from imposing a device requirement that is different from or additional to a federal one — and the Supreme Court has tied that bar to the pathway, not the device. PMA approval preempts state tort claims. 510(k) clearance, as a rule, doesn't. De Novo is still being litigated.

By Connor Griggs — Regulatory & Quality Strategist

Two devices can do almost the same clinical job, clear FDA review in the same year, and end up on opposite sides of a line that has nothing to do with either device’s safety record: one manufacturer can have a state tort claim thrown out before discovery starts, and the other cannot, because of which premarket pathway the device went through. That line is federal preemption, and the statute that draws it reads the same for every device. The courts have not applied it the same way to every pathway.

The statute, and what it actually bars

21 U.S.C. § 360k(a) is short: no state may establish or continue in effect, with respect to a device, a requirement that is both different from, or in addition to a federal requirement applicable to that device, and that relates to safety or effectiveness or another matter the federal requirement covers. Section (b) lets a state petition FDA for an exemption from that bar; it is rarely used and nothing turns on it here. What matters for every device company is what counts as a federal “requirement” specific enough to trigger the first clause — and the Supreme Court has answered that question twice, with two different results, for two different pathways.

Lohr: 510(k) asks the wrong question to count

In Medtronic, Inc. v. Lohr (1996), the Court held that a 510(k) clearance does not impose a device-specific federal requirement at all. Section 510(k) review asks whether a device is substantially equivalent to a device already on the market before 1976 — it does not ask FDA to independently weigh that device’s own safety and effectiveness the way a PMA review does. A process built around resemblance, not an independent finding, leaves nothing specific enough to preempt. The result: a state tort claim over a 510(k)-cleared device’s design or warnings is not, as a rule, barred by § 360k(a).

Riegel: PMA approval is the requirement

Riegel v. Medtronic, Inc. (2008) drew the opposite line for the other pathway. A PMA review weighs a specific device’s own safety and effectiveness against FDA’s own standards before approval — and that approval, with its conditions, is itself a federal requirement specific to that device. A state claim asking a jury to hold the device to a different design, different labeling, or a different manufacturing standard than what FDA approved is a requirement “different from, or in addition to” the federal one. Riegel held it preempted.

The Court left one door open even for PMA devices: preemption, it wrote, “does not prevent a State from providing a damages remedy for claims premised on a violation of FDA regulations” — a claim that parallels the federal requirement rather than adding to it. Riegel itself did not have to decide how far that door opens; lower courts have split since on how closely a state claim has to track an actual federal violation to count as parallel rather than additional.

De Novo: no Supreme Court answer yet

Neither Lohr nor Riegel addressed the pathway that has grown the most since both were decided. De Novo classification sits between the two in the regulatory scheme — it reaches a device with no predicate, grounded in FDA-issued special controls, without PMA’s full premarket-approval review. For most of three decades, courts read Lohr broadly enough to treat any device outside the PMA track, De Novo included, as unpreempted by default. In July 2024, a federal district court in Louisiana broke from that default: in Dickson v. Dexcom, Inc., the court held that § 360k(a) preempted safety and effectiveness claims against a Class II continuous glucose monitor cleared through De Novo, reasoning that the device-specific special controls FDA built into its own classification order were themselves a federal requirement under Riegel’s test — not a 510(k)-style resemblance finding under Lohr’s. It is one district court decision, not an appellate holding, and not binding outside its own docket, but other courts have since cited it for the same proposition.

The practice

None of this changes what a classification order, a clearance letter, or a PMA approval says about a device’s market status — openFDA’s own clearance and classification records carry no field for preemption at all, because preemption is not a determination FDA makes; it is a consequence courts draw from the record FDA already made. A pathway decision that looks purely like a cost-and-timeline tradeoff on a regulatory strategy memo carries a second, separate consequence that shows up only if the device is ever sued, decided by a different body, under a different body of law, years after the clearance letter is filed away. Whether a specific product’s own facts fall on one side of that line or the other is a question for litigation counsel working that product’s own record — this is a description of how the doctrine is built and where it currently stops, regulatory intelligence and method, never regulatory or legal advice about any specific device’s litigation exposure.

Practice
A provider letter doesn't need a recall
Method
A qualified tool has a context, not a blanket
Practice
The eMDR account isn't automatic
Method
The manufacturer who owns no factory
Practice
GLP covers the safety study, not the device
Method
Part 11 governs the copy, not the duty.
Method
A K-number isn't always a 510(k)
Practice
Approved once. Reported every year.
Method
The goal date passes. FDA owes a letter.
Method
The money a 510(k) never has to mention
Method
A discussion paper isn't guidance
Practice
A stereotaxic system isn't a RASD
Method
The A-list is a forecast, not a deadline.
Practice
Ultra has a gram range. Fragrance doesn't.
Method
Patient preference just reached the 510(k)
Practice
The health hazard evaluation isn't the record
Method
The sponsor proposes risk. The IRB decides.
Practice
A Special 510(k) needs a predicate you own
Method
The flag marks the reviewer, not the result
Practice
Call it fifteen. File it in thirty.
Method
The recognition number isn't the list number
Practice
The letter to file isn't in the guidance
Practice
Cleared doesn't code. Coded doesn't cover.
Practice
Exempt from 510(k). Restricted anyway.
Method
Exempt from registration. Not from 510(k).
Method
Right to try reaches drugs, not devices
Practice
The 510(k) clears the range, not the fit
Practice
The hospital's duty runs to the maker, not FDA.
Method
The predicate study isn't the waiver study.
Method
No clinical hold for an IDE. Only disapproval.
Practice
A CFG can cover a device that never came here.
Method
The symbol skips text. The glossary can't.
Practice
The label clears 812.5. The flyer doesn't.
Method
A 510(k) can change hands. No rule says how.
Practice
A missed renewal is two violations, not one
Method
Drugs expire by rule. Devices by choice.
Practice
Denied the petition. Not the outcome.
Practice
Starting a committee is public. Renewing isn't.
Method
A recall closes. A radiation refund doesn't.
Method
The trial is public. The IDE isn't.
Practice
A PMA withdrawal pays back. A 510(k) doesn't.
Method
Interoperable is a design term, not a class.
Practice
The category is new. The 510(k) record isn't.
Practice
The device is cleared. Not the room.
Method
Classified in 2011. Excluded in 2016.
Practice
The law reached devices. The guidance hasn't.
Method
The label says RUO. FDA reads intent.
Method
Humanitarian. Not always nonprofit.
Practice
The certification has a one-year tail
Method
The code changed. The failure didn't.
Practice
Skip has a ceiling. Search after doesn't.
Method
Reviewed together. Decided separately.
Practice
Class III. Not called for PMA.
Practice
The standard for the wire, not the device
Method
A reference device isn't a predicate
Practice
A predicate checklist nobody has to follow
Method
The pathway that skips the predicate
Method
A network FDA funds, not FDA runs
Practice
Panel-recommended. Not yet classified.
Method
The pilot that needed a statute
Practice
A committee no clearance ever names
Method
Never cleared here. Exported anyway.
Practice
Built for drugs. Silent on devices.
Practice
A significant decision carries its own clock
Method
The guidance that went back to draft
Method
Breakthrough crosses centers. TAP does not.
Practice
Two lists decide who reviews your 510(k)
Method
The reprocessor holds the 510(k)
Practice
The Rx symbol never reaches the record
Method
The disqualification list names no device
Practice
The summary FDA writes, not the sponsor
Practice
A missed MDR is misbranding, not a defect
Method
Two names on one registration, two jobs
Practice
A K-number is public. The other isn't.
Method
Exempt covers the type, not the use
Method
Recognized. Not published.
Practice
The supplement got eSTAR. The PMA did not.
Method
Congress ordered a guidance, not a rule
Practice
AI-enabled is a list. Not a field.
Practice
The companion diagnostic list changed its name
Method
The master file neither of you can read
Practice
The condition inside the approval order
Method
The repair FDA calls manufacturing
Method
Vacated in March. Filed in September.
Practice
Classified. Not waived.
Practice
Emergency use isn't defined in Part 812
Method
The two-year inspection clock is gone
Method
The PMA clock starts at the last module
Practice
The Q-Sub that doesn't buy you more time
Method
Lapsed and revoked end an EUA differently
Practice
A recall strategy has a grade. openFDA doesn't.
Method
A granted De Novo becomes a predicate.
Practice
Least burdensome changes the ask, not the bar.
Method
The lab is accredited. Not the device.
Practice
Breakthrough is law. STeP is a guidance.
Practice
MDSAP replaces routine. Not for-cause.
Method
No field on a 510(k) says what it treats
Method
Discretion is a policy. Exclusion is a statute.
Practice
Designated by one office. Approved by another.
Method
MedSun sits beside MAUDE, not inside it
Practice
The five-unit device never gets a K-number
Practice
Meeting the criteria isn't a tracking order
Method
The PMA clock doesn't stop at approval
Practice
The K-number doesn't track today's device
Method
A MAUDE report proves it was filed
Practice
Not every MDR starts with the manufacturer
Method
Completed. Not terminated.
Practice
The device-ban list has three entries
Method
The UDI exception has its own deadline
Method
MDUFA VI is being negotiated in public
Practice
The inspection manual changed, not the rule
Method
The count Congress had to legislate
Practice
The small-business rate isn't retroactive
Method
A detention order has no docket to read
Practice
PMA approvals moved off the Federal Register
Method
A citizen petition names names
Practice
A market withdrawal isn't a recall
Method
Not every device letter comes from CDRH
Practice
Closed on one device. Open on the rest.
Method
No IDE is confirmed until the PMA is
Practice
The remedial-action call sets the clock
Method
Jurisdiction is a letter, not a field
Practice
An address is not a footprint
Method
The final rule that isn't final yet
Practice
The special controls aren't in the record
Method
Exempt from GMP. Not from complaints.
Practice
The notify list isn't the shortage list
Method
A constraint can't be skipped. A script can.
Practice
No cron sends this. A person does.
Method
Corrected, not rewritten
Practice
Quiet is a finding, not a default
Method
The summary is public. The statement isn't.
Practice
A competitor's 513(g) leaves no trace
Practice
An empty response field proves nothing
Method
A petition is public. Its outcome isn't.
Practice
The panel is a category, not a meeting
Method
An accessory carries its own classification
Method
A TPLC report is a category, not a device
Practice
A 522 order is a question, not a verdict
Method
A PCCP change leaves no new record
Practice
Not every signal becomes a recall
Method
eSTAR did not remove the acceptance check
Practice
An early alert has no recall number yet
Practice
A consent decree isn't a database row
Method
The record has no cybersecurity field
Practice
Classification Change is not one event
Method
Publication date is not effective date
Method
One event. Many recall numbers.
Practice
Quantity is a sentence, not a number
Practice
How a firm told you is a field too
Method
A RIN outlives the document number
Method
The docket ID is not the comments
Practice
The 510(k) record doesn't name the predicate
Practice
The 510(k) FDA didn't review first
Method
The letter FDA doesn't have to post
Practice
Cleared doesn't say which track it took
Method
The clock stops. The calendar doesn't.
Method
Why a guessed match never reaches you
Practice
The standards list never names a device
Method
The guidance isn't in the Federal Register.
Practice
Filtered. Not discarded.
Method
There are two recall databases, not one
Practice
Breakthrough status is confidential by law
Practice
One MAUDE report is not one malfunction
Method
A supplement number is not a track
Method
Cleared is not approved
Practice
Nationwide doesn't mean your building
Practice
Not every correction reaches FDA
Method
Registered. Not reviewed.
Method
Contact ODE is not a pathway
Practice
Almost no recall is FDA-mandated
Method
FDA's device names read like a card catalog
Practice
openFDA is free. It is not unlimited.
Method
GUDID stores the model. Not the box.
Practice
Detention doesn't need a warning letter
Method
A Class I device, a Class I recall
Practice
21 CFR 820 didn't move. Its contents did.
Method
The product code that doesn't exist yet
Practice
The classification posts. The 483 behind it doesn't.
Method
A recall has three dates, and the pipeline had to pick one
Practice
The count is real. The rate is not.
Method
The firm on the record is not the firm on the box
Practice
Low confidence is an instruction, not a hedge
Method
The same company, spelled three ways
Practice
A device that was never a medical device
Method
FDA's warning letters, addressed by column number
Practice
Your regulation has a decimal. FDA's watch doesn't.
Method
Three letters is too short to search for
Practice
Most warning letters never close
Method
The guidance that skipped the draft
Practice
Ongoing, as of when?
Method
The least interesting fact in a 510(k)
Practice
No recall arrives with a product code attached
Practice
The deadline that doesn't email you
Method
The warning letter has two dates
Method
How to monitor FDA without drowning
Practice
Your predicate was recalled. Now what?
Method
Why no item reaches you without a human